Sammanfattning
Over the last decade Fourier transform infrared (FT-IR) and near-infrared (NIR) spectroscopic imaging with focal plane array (FPA) detectors have proved powerful techniques for the rapid visualization of samples by a combination of spectroscopic and spatial information. Using these methods, selected sample areas can be analyzed with reference to the identification and localization of chemical species by FT-IR spectroscopy in the transmission or attenuated total reflection (ATR) mode and by NIR spectroscopy in diffuse reflection with a lateral resolution in the micrometer range. The present communication focuses on the quantitative determination of the active ingredient composition of a solid drug formulation by NIR spectroscopic imaging with a focal plane array detector and the results obtained are compared to the quantitative data obtained by conservative light-fiber NIR spectroscopic diffuse reflection measurements with a single-element detector. The communication also addresses the issue of penetration depth of NIR radiation into the investigated solid material.
| Originalspråk | Engelska |
|---|---|
| Sidor (från-till) | 1200-1208 |
| Antal sidor | 9 |
| Tidskrift | Applied Spectroscopy |
| Volym | 62 |
| Nummer | 11 |
| DOI | |
| Status | Publicerad - 2008 |
Nyckelord
- Near-infrared spectroscopy
- NIR spectroscopic imaging
- Focal plane array detector
- FPA
- Diffuse reflection
- Solid drug formulation
- Active pharmaceutical ingredient
- API