Sammanfattning
Even though size exclusion chromatography (SEC) with post column addition of calcofluor (SEC-calcofluor) has been used for the determination of cereal beta-glucan molar mass in foods for many years, there is a lack of systematic evaluation of the method. To address this issue a set of suitable beta-glucan standards were generated by preparative SEC and their molar mass characteristics were determined by analytical multi-detection SEC (refractive index (RI), light scattering). Each standard was then analysed by SEC-calcofluor at three different labs. As a direct comparison, the analyses were repeated with a RI detector. For SEC-calcofluor accurate measurements of weight average molar mass (M-w) can be made for beta-glucan populations within 10-500 x 10(3) g/mol. Above this molar mass threshold there is an increasing tendency for underestimation of M-w. Precipitation of some beta-glucan-calcofluor complexes may have delayed their transport into the detector. (C) 2015 Elsevier Ltd. All rights reserved.
| Originalspråk | Engelska |
|---|---|
| Sidor (från-till) | 254-264 |
| Antal sidor | 11 |
| Tidskrift | Carbohydrate Polymers |
| Volym | 124 |
| DOI | |
| Status | Publicerad - 2015 |
Nyckelord
- Cereal beta-glucan
- Calcofluor
- Molar mass
- Chromatography
- Light scattering
Citera det här
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver