Återgå till huvudnavigering Återgå till sök Gå direkt till huvudinnehållet

Geometrical method for interpolating S-peaks from cow ECG using a microcontroller

  • M. Johannes Tiusanen
  • , Mikko I. Hautala
  • , Emma Ternman
  • , Matti Pastell

    Publikation: Bidrag till tidskriftArtikel i vetenskaplig tidskriftPeer review

    Sammanfattning

    This paper describes a new and computationally efficient adaptive geometrical method for improving the resolution of electrocardiogram (ECG) S-peak detection. The target device is an ECG implant for dairy cows which provides limited calculation power. The method utilises only arithmetic operations and is adaptive to individual ECG curves and electrode installation effects. The method tracks the average heights of the sample points preceding and following the S-wave peak. Comparing these heights to the long-time average values provides an estimate of the offset of the peak sample from the actual peak moment. On average, the prediction of spline interpolation on the original 512 Hz turning point is 0.043 ms more accurate than our geometrical algorithm. Both methods result in less than a 0.7 ms difference from the original data. However, the proposed method is significantly more computationally efficient than cubic spline interpolation. (C) 2014 IAgrE. Published by Elsevier Ltd. All rights reserved.
    OriginalspråkEngelska
    Sidor (från-till)324-328
    Antal sidor5
    TidskriftBiosystems Engineering
    Volym129
    DOI
    StatusPublicerad - 2015

    Nyckelord

    • ECG
    • Interpolation
    • Cow
    • S-wave

    Citera det här