Anterior chamber depth estimation by Scheimpflug photography

    Publication: Contribution to journal › Journal article › peer-review

    Abstract

    Purpose: To obtain a model suitable for evaluation of the anterior chamber depth (ACD) maintaining capacity of viscoelastic substances - a critical parameter in cataract surgery.Methods: ACD was estimated using an anterior segment analysis system (EAS-1000, Nidek Co Ltd, Japan). Variation was studied as an effect of batch number and eye within batch, storage, globe fixation, globe position, and repeated analyses of Scheimpflug images.Results: Considerable variation in ACD was found between eyes, batches, and as a result of prolonged storage (biological factors). Different IOPs before surgery caused significant alterations in ACD after overfilling with a viscoelastic substance. No significant differences were found when the globe position was altered or when the ACD was re-estimated repeatedly from the same photograph.Conclusion: Differences due to biological factors are more important than methodological errors caused by the Scheimpflug instrument. To reduce these variations when viscoelastics are evaluated, all measurements should be calculated as % of ACD before surgery and the anterior chamber should be overfilled with a viscoelastic solution.
    Original languageEnglish
    Pages (from-to)576-579
    Number of pages4
    JournalActa Ophthalmologica Scandinavica
    Volume79
    Issue number6
    DOIs
    Publication statusPublished - 2001

    Keywords

    • anterior chamber depth
    • ACD
    • viscoelastic solution
    • Healon
    • EAS-1000 Scheimpflug
    • porcine eyes