Abstract
Purpose: To obtain a model suitable for evaluation of the anterior chamber depth (ACD) maintaining capacity of viscoelastic substances - a critical parameter in cataract surgery.Methods: ACD was estimated using an anterior segment analysis system (EAS-1000, Nidek Co Ltd, Japan). Variation was studied as an effect of batch number and eye within batch, storage, globe fixation, globe position, and repeated analyses of Scheimpflug images.Results: Considerable variation in ACD was found between eyes, batches, and as a result of prolonged storage (biological factors). Different IOPs before surgery caused significant alterations in ACD after overfilling with a viscoelastic substance. No significant differences were found when the globe position was altered or when the ACD was re-estimated repeatedly from the same photograph.Conclusion: Differences due to biological factors are more important than methodological errors caused by the Scheimpflug instrument. To reduce these variations when viscoelastics are evaluated, all measurements should be calculated as % of ACD before surgery and the anterior chamber should be overfilled with a viscoelastic solution.
| Original language | English |
|---|---|
| Pages (from-to) | 576-579 |
| Number of pages | 4 |
| Journal | Acta Ophthalmologica Scandinavica |
| Volume | 79 |
| Issue number | 6 |
| DOIs | |
| Publication status | Published - 2001 |
Keywords
- anterior chamber depth
- ACD
- viscoelastic solution
- Healon
- EAS-1000 Scheimpflug
- porcine eyes